Standard setting – Flowserve ORGS 11-2 User Manual

Page 13

Advertising
background image

13

The measuring range can be switch-selected between

≥

0.5 µS/cm and

≥

10 µS/cm

by means of a code switch :

1. Undo screws of the electrode terminal and remove housing cover , fig. 3.

Code switch 1 - 4 OFF

Measuring range

≥

0.5 µS/cm

Code switch 1 - 4 ON

Measuring range

≥

10 µS/cm

2. Mount housing cover and fasten screws .

Standard Setting

Factory setting

The equipment features the following factory set default value:

■

Measuring range

≥

10 µS/cm

Switch selection of measuring range

Attention

■

Screwdriver for cross head screws, size 1

■

Screwdriver for slotted screws, size 2.5; completely insulated according to
VDE 0680

■

Make sure not to damage the electronic components when setting
the code switch.

■

Do not use a pencil to set the code switch!

Tools

H

J

H

J

5