9

Technical Data

– continued –

Max. admissible ambient temperature
0 °C to +70 °C

Case
19" slide-in unit with front panel to DIN 41494 part 5 and rear 32 way Euro card
connector to DIN 41612 installed in a mounting panel or for installation in
19" mounting panel.
Front panel, mounting panel: aluminium

Wiring
Design “b”:
screw-type terminal strips at the back of the mounting panel, max.
conductor size 1.5 mm²
Design “c”: 32 pole screw-type connector at the back of the 19" mounting panel, max.
conductor size 1.5 mm²
Supply line to level transmitter: screened two-core cable, conductor size 0.5 mm²,
max. cable length 150 m

Internal fuse
Glass cartridge fine-wire fuse M 0.05 A, 5 x 20, replaceable

Weight
approx. 0.8 kg

Explanatory Notes