12 troubleshooting, 13 replacement parts – Fluke Biomedical 06-526 User Manual

Page 24

Advertising
background image

Nuclear Associates 06-526 & 06-526-2200
Operators Manual

3-4

8. Measure 5 exposures. Record the average as "Measured." All five should be within 3%.

9. Calculate correction factor (cf):

cf = True

Measured

10. Reposition RAD-CHECK PLUS and measure 1 exposure to verify that the x-ray machine did not

change.


All readings on the RAD-CHECK PLUS using the external chamber should be multiplied by the correction
factor. (HINT: Write the correction factor on a small sticker and put it on the chamber.)

3.12 Troubleshooting

Extreme caution should be used when making
connections with the chambers and rear panel
connectors.

An electrical shock hazard exists between the ion
chamber bias connector and ground (HV Out).


When there is a problem with the unit, refer to the table below for possible causes and corrective action.

Symptom Possible

Cause

Correction

Action

No display &
No reset light

Dead battery
Broken battery snap lead
Defective ON/OFF switch

Replace battery
Replace lead
Replace switch

Display ON;
No reading

Rear panel selector switch positioned
incorrectly

Check and correct switch position

Inaccurate readings Exposure level too low (see Figure 2-3) Position rear panel selector switch to

MAN reset position

Does not zero

Adjust rear panel zero (see Zero Set)

Low battery voltage

Replace battery

Defective/damaged ion chamber

Recalibrate chamber
Return unit to factory for chamber
repair/replacement


3.13 Replacement Parts

Unless otherwise noted, resistors are .25 W, 5% carbon film; non-polarized capacitors are ceramic disk
type, 20% tolerance, 100 VDC; and capacitors marked "FILM" are 10% tolerance, 100 VDC. Non-limited
parts are standard components and should be available through local suppliers.

If it is necessary to order replacement parts, contact Fluke Biomedical at 440.248.9300.

WARNING

WARNING

Advertising
This manual is related to the following products: