Nuclear Associates 06-526 & 06-526-2200
Operators Manual

1-2

1.2 Specifications

RAD-CHECK PLUS

Range Dose

0.001 to 1.999 R (.01 to 19.99 mGy)

Range Rate

0.01 to 19.99 R/min (0.1 to 199.9 mGy/min)

Measurement Area

18.3 cm

2

(4.8 in Ø) effective

Standard Calibration

5% at 75 kVp with 4 mm Al filtration at 22° C (72° F) and one
atmosphere - 760 mm of Hg

Chamber Bias Supply 270 V

Reproducibility

Short term within 1% of full scale

Energy Response

5% from 30 kVp to 150 kVp

Electrometer Drift

< 1 mR/min

Maximum
Instantaneous
Exposure Rate

20 R/sec

Reset Switch

±1 mR of zero

Power Supply

9 V alkaline battery (Duracell type MN 1604 or equivalent). 50 hour
operation in manual or external mode; 100 hour operation in auto
mode

Display

3 1/2 digit LCD; 1/2 inch high digits; low battery indication

Operating Conditions

10° to 40° C (50° to 104° F)

Relative Humidity

90%, non-condensing

Dimensions

6 (w) x 6.3 (d) x 2.5 in (h) (154 x 160 x 64 mm)

Weight

1.1 lbs (500 g)

Optional External Detectors (Part No. 6000-528 (30 cc) & 6000-529 (3 cc Mammo))

Radiation

Model 6000-528: x-rays from 30 to 150 kVp

Measurement

Model 6000-529: x-rays from 16 to 90 kVp

Nominal Sensitivity

Model 6000-528: 9 nC/R

Model 6000-529: 1 nC/R

Volume

Model 6000-528: nominal 30 cc

Model 6000-529: nominal 3 cc

Energy Response

Model 6000-528: 30 to 150 kVp, within 7%

Model 6000-529: 16 to 90 kVp, within 7%

Wall

0.031 inch phenolic

Termination

Signal; coaxial, fully guarded BNC HV; recessed banana plug

Cable Length

15 feet

Dimensions

Model 6000-528:
4 (w) x .54 (d) x 4 in (h) (101.6 x 13.7 x 101.6 mm)

Model 6000-529:
1.57 inch Ø x .54 inch deep (40 mm x 15 mm)

* Specifications are subject to change without notice.