Fluke Biomedical 35080M User Manual
Page 10

General Information
Specifications
1
1-3
Table 1-2.
The Wide Range Filter Pack (37617) Specifications
Range: 50 to 150 kVp
Sensitivity: 0.1 V/10 kV
Cal: 0.400 V
Zero: 40 kV
Cal: 80 kV
Accuracy: ± 2% exclusive of linearity and filtration
effects
Linearity and filtration effects can be corrected using
curves from Section 2 of this instruction manual.
Optional Filter Packs
A K-edge Mammo Filter Pack (37355)* and a Linear Mammo Filter Pack (37351) used for precision
measurements from 22 to 40 kVp, a CT Filter Pack (33551) used for CT system kVp measurements from
70 to 140 kVp, a Mobile Filter Pack Plus (37946) used for measurements on X-ray systems with kV
waveforms above 1 kHz with substantial ripple, and a Low Range Filter Pack (38237) used for
measurements from 30 kVp to 90 kVp where the kVp range or sensitivity of other filter packs is not
adequate.
* Patent numbers 4,843,619 and 4,916,727