Linear Mammo Filter Pack Option (37351)

Corrections

5


5-3




Figure 5-2. The correction curve that shows the effect of adding 0.25mm Al (0.01in Al) to the 24 microns

molybdenum filter. The aluminum makes the Linear Mammo Filter Pack (37351) read high.
To make corrections, subtract the kV correction from the indicated reading.


In most cases, amount of molybdenum filtration is unknown and the use of the Cadmium K-edge Mammo
Filter Pack (37355) for precise calibration of the generator at a kVp between 27.5 and 29.5 kV is required.



See Appendix A for more detailed instructions
on the use of the K-edge and Linear Mammo
Filter Packs (37355 and 37351 respectively).

5.5 Operating Instructions

This procedure will normally follow that given in Section 5.4 where measurement with the Cadmium K-
edge Mammo Filter Pack (37355) ended.

1. Measurement

Setup

- Replace the Cadmium K-edge Mammo Filter Pack (37355) with the Linear

Mammo Filter Pack (37351). Reposition the Model 35080M kVp Divider to approximately 15 inches
from the focal spot by lowering the compression paddle or other support device (see Figure 5-3).

NOTE

Linear Mammo Filter Pack (37351)

Correction for Aluminum

kV

Correct

ion

Indicated kVp