2 specifications, 3 filtration effect corrections, Range – Fluke Biomedical 35080M User Manual

Page 41: Sensitivity, Zero, Calibration filtration, Example

Advertising
background image

35080M
Operators Manual

7-2

7.2 Specifications

Range

50 to 135 kV

Sensitivity

0.1 V/10 kV

Cal

0.600 V

Zero

40 kV

Cal

100 kV

Calibration Filtration

2.5 mm Al

7.3 Filtration Effect Corrections

Corrections for filtration are made using the correction curve shown in Figure 7-1. The Mobile Filter Pack
Plus (37946) is calibrated at 2.5 mm Al total filtration, so measurements made at other filtration values
require correction of the measured reading. Corrections are made by subtracting the total filtration of the
X-ray machine being measured from 2.5 mm Al, multiplying the result by the value read off the chart for
the measured kVp, and adding this result to the measured kVp.

Example

The Model 35080M kVp Divider with a Mobile Filter Pack Plus (37946) inserted reads 80kVp. The X-
ray system has 3.3 mm Al total filtration. The filtration effect curve of Figure 7-1 indicates a filtration
effect of 1.05 kVp per mm Al at 80 kVp. The difference in filtration, 2.5 - 3.3 = -0.8 mm Al, is multiplied
by the correction value to yield -0.8 x 1.05 = -0.84 kVp. This number is then added to the apparent
reading to obtain the actual kVp, which is 80 kVp + (-0.84 kVp) = 79.16 kVp.



















Figure 7-1.

Mobile Filter Pack Plus (37946) Filtration Effect Correction Curve

Filtration Effect

kVp Per mm Al

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

40

50

60

70

80

90

100

110

120

130

140

Model 35080B Reading kVp

k

V

p

Pe

r

mm

Al

Advertising