Fluke Biomedical 35080M User Manual
Page 15
![background image](/files/785261/content/doc015.png)
35080M
Operators Manual
2-4
In positioning the kVp divider, the long axis of the instrument should be perpendicular to the X-ray tube
axis to eliminate “heel effect.” The beam must strike the filter pack side of the kVp divider; the unit is
designed to be turned over for under-table X-ray heads. In setting the X-ray generator, 200 mA at 22
inches is recommended, but the limits of operation allow higher or lower settings and distances. Refer to
Section 1-4, Specifications, for low current limits.
After correctly positioning the kVp Divider, connect the oscilloscope, place the kVp Divider in the “RUN”
mode and make an exposure.
ABOVE TABLE
X-RAY TUBE
TUBE AXIS
LONG AXIS OF 35080M
X-RAY TABLE
TUBE AXIS
UNDER TABLE X-RAY TUBE
X-RAY TABLE
LONG AXIS OF 35080M
Figure 2-2. Geometry of Setup for Above-Table
Figure 2-3. Geometry of Setup for Below-Table
X-ray
Tube
X-ray
Tube