Section 3, Calibration, 1 calibration – Fluke Biomedical ASM-990 Series User Manual

Page 25

Advertising
background image

Calibration

Calibration

3

3-1

Section 3

Calibration

3.1 Calibration

Probe calibration parameters (probe profiles) are NOT operator accessible through the front panel.
Rather, they will be downloaded, through the usage of PC based calibration software, via the IrDA port.

Parameters for each calibrated probe are then stored/saved in the unit’s non-volatile memory (EEPROM).
The user may then select from one of the previously stored probe data calibration parameters. However,
the user then cannot select units (e.g. mR, uCi, nSv, etc.) unless that particular probe has been calibrated
to operate with those specific units.

If no previously stored probe data has been selected, then the unit will default to the non-calibrated
“CPM,” or “COUNTS PER MINUTE,” mode only.