Product specifications – Ivoclar Vivadent Quick User Manual

Page 16

Advertising
background image

16

9. Product Specifications

9.1

Delivery form

1 Quick
1 Anti-glare shield
1 Power cord
1 Extra fuse
1 Operating Instructions
1 Warranty card

Colour:
White

(RAL 9016)

9.2

Technical data

Power supply
Single-phase alternating current

230, 240 V / 50-60 Hz

100, 110, 120 V / 50-60 Hz

Tolerated voltage fluctuations

± 10%

Power consumption

max 100 W

Voltage of lamp

12 V

Type of lamp

Halogen lamp for Quick (only use lamps from Ivoclar Vivadent)

Electrical fuses
Values:

T 800 mA (230/240 V) has to comply with IEC 127

T 1.6 A (100/110/120 V) has to comply with UL and CSA

Dimensions of fuses

230-240 V = ø 5 x 20 mm
100-120 V = ø 5 x 20 mm

Dimensions of unit

Length: 390 mm;

Width: 150 mm;

Height: 130 mm (closed)

Lightwave range (filtered)

400 - 500 nm

Weight

2.3 kg

Safety information
The Quick complies with the following guidelines:

IEC 1010-1/EN 61010, Part 1

UL 3101-1

CSA 22.2 No. 1010-1

Directive 89/336/EEC, radio protection / electromagnetic
compatibility

EMC tested

9.3 Acceptable operating conditions

Acceptable temperature range
5–35 °C (41–95 °F)

Acceptable humidity range
Maximum relative humidity at 31 °C (87 °F) gradually
decreasing to 50 % at 35 °C (95 °F), condensation excluded

Acceptable ambient pressure
The unit may be used at altitudes of up to 2000 m
above sea level.

Max. operation
60 seconds

9.4

Acceptable transportation and storage
conditions

Acceptable temperature range
-20 to +50 °C (-4 to +122 °F)

Acceptable humidity range
Maximum relative humidity at 31 °C (87 °F) gradually
decreasing to 50 % at 40 °C (104 °F), condensation excluded

Acceptable ambient pressure
500-1060 mbar

Use only original packaging together with the respective foam
material for shipping purposes.

Quick-BA-5spr.qxd 9.8.2006 7:28 Uhr Seite 16

Advertising