8. What If ...?

8.1 Technical malfunctions

8.2 Repairs

Repairs may only be carried out by the personnel of a certified

Ivoclar Vivadent Service Center. Please contact your Customer
Service representative.

Malfunction

1 The pump is not working

2 The vacuum performance is

declining or is bad

3 Vacuum pump is unusually noisy

Double-check

– Is the pump correctly connected?

– Has the thermoswitch been

activated because of overheating?

– Sound absorber clogged

– Has the test program for the

pump been carried out?

– Is the sound absorber defective or

incorrectly mounted?

– Is there a leak in the furnace sys-

tem or the pump?

Action

– Connect pump correctly.

– Allow the pump to cool and

switch on again.

– Check pump location for

adequate circulation of air.

– If the problem cannot be solved,

contact your local Ivoclar Vivadent
Service Center.

– Replace sound absorber.

– Carry out the test program for the

pump (PX1).

– Send vacuum pump to Service

Center for maintenance.

– Connect sound absorber correctly

or replace it, if it is defective.

– Check connections and hose for

tightness.

– Clean sealing of the furnace head.

– Check pump performance with

the corresponding test program.