3 installation of pr-23 pharma refractometer, 1 hardware and software requirements, 2 mechanical and electrical requirements – K-Patents PR-23 for K-Patents Appendix User Manual
Page 15: Installation of pr-23 pharma refractometer

© Copyright K-Patents 2013. All rights reserved.
15
3
Installation of PR-23 Pharma
Refractometer
3.1 Hardware and Software Requirements
PR-23 software is included in the Indicating Transmitter DTR and it comprises the
following functions:
● Automatic temperature compensation
● Ethernet connection for data download
● Sensor diagnostics and verification
● The K-Patents Computer Communicaton software CC-11 is an optional
communication product.
3.2 Mechanical and Electrical Requirements
Power supply for the refractometer is AC input 100-240 VAC/50-60 Hz, optional 24 VDC,
30 VA.
3.3 Sensor installation for use in Pharmaceutical Batch
Manufacturing
Laboratory table top or trolley installation and key considerations for the site
preparation
1. Physical dimensions of the instrument and accessories: make sure there is enough
space to accommodate them.
2. Suitable recommended operational environment for the instrument and for the Cargille
Refractive Index Liquids should be maintained between 20 – 30 °C (68 – 86°F).
3. Utilities: 100-240 VAC/50-60 Hz (optional 24 VDC, 30 VA) electrical power supply and
computer network connection.