Typical faults and causes – K-Patents PR-23 Troubleshooting Guide User Manual

Page 11

Advertising
background image

TROUBLESHOOTING GUIDE
FOR PR-23 ALL MODELS

11 (14)

January 9, 2015

K-PATENTS OY

Postal Address: P.O.Box 77, FI-01511 Vantaa, Finland

Street Address: Elannontie 5, FI-01510 Vantaa, Finland

Tel. int.+358 207 291 570

Fax int.+358 207 291 577

[email protected]

www.kpatents.com

Vat No. FI03035575

Business ID 0303557-5

Registered in Helsinki

4. Typical Faults and Causes

4.1 Description: Measurement is constantly showing 0 concentration

or “flat line” in DCS

Cause(s): Error message on the measurement status, such as No optical image, no sensor, short circuit, no

sample etc. During certain error messages

the instrument gives “default” mA-output, this is configured to

give 4 mA as preset.

Solution: Troubleshoot the issue indicated by the error message.

4.2 Description: Measurement Status No optical image, low image quality

Cause(s): The CCD-element does not receive enough light to determine the borderline of reflected light vs

shadow on the optical image. Lack of light in 90 % of the cases is caused by coated or eroded prism. Liquid

leakage through the prism gasket also fits this category. Remaining 10 % of the faults are related to the other

optical or electrical faults, such as broken optical fiber or detached flat cable. Severely bad optical images

are generally seen with these error messages.

Solution:

Check the functionality of the washing system with manual washes. If that won’t help, clean the

prism with a glassfiber brush and check the optical image. Also check the prism for scratches or marks. The

prism needs to be replaced, if there are any visible marks on the polished surfaces. Possible leakage can be

detected from abnormal temperature readings and increased head humidity. Services can be done only by

K-Patents or a trained service person.

4.3 Description: Spiking concentration outputs

Cause(s): Coating of the prism or prism erosion causes a low quality optical image and unstable

measurement.

Solution: Clean the prism. If the prism is eroded, service accordingly. If the coating is a constant problem,

consider other installation position or additional washing system. Check the functionality of the washing

system, if there already is installed one.

4.4 Description: No connection to

the sensor “No Signal”, “No Sensor” or endless “Starting up..”

status

Cause(s): Weak cable connections, faulty sensor processor card, or electrical component failure due to

excessive humidity or temperature inside the sensor. Another potential cause may be a liquid leakage inside