K-Patents PR-23 Troubleshooting Guide User Manual

Page 7

Advertising
background image

TROUBLESHOOTING GUIDE
FOR PR-23 ALL MODELS

7 (14)

January 9, 2015

K-PATENTS OY

Postal Address: P.O.Box 77, FI-01511 Vantaa, Finland

Street Address: Elannontie 5, FI-01510 Vantaa, Finland

Tel. int.+358 207 291 570

Fax int.+358 207 291 577

[email protected]

www.kpatents.com

Vat No. FI03035575

Business ID 0303557-5

Registered in Helsinki

2.2

Interpretation of the Diagnostic Values

The instrument has several diagnostic values that can be used to inspect the condition of the instrument and

for troubleshooting purposes. Especially, the LED and the QF-value can be informative when interpreting the

optical images.

Table 1

Instrument diagnostic values

Explanation

Units / Range

CONC

Concentration

several units

TEMP

Temperature

°C / °F

CCD

Light activation of the CCD

%

nD

Refractive index

-

CALC

Calculated concentration

raw value

QF

Quality factor

typically 0-200

LED

Amount of light from the LED 0-100 %

HD TEMP Inside temperature

alarm at 65 °C

HD HUM

Inside humidity

alarm at 60 %

Measurement status: Indicates the current status of the measurement. It also displays error messages. The

messages can be as follows: Normal operation, No sample, Prism wash, Recovering, No sensor, High

sensor humidity, etc. The complete list of different messages with descriptions can be found in the instruction

manual.

TEMP = Temperature element of the instrument is located on the tip of the sensor probe. The temperature

element is one of the first components affected by prism gasket leakage. Typically the leakage is detected

from abnormal temperature readings as the resistance of the PT-1000 temperature element increases when

it is corroded. If the electric circuit is cut off, the temperature will drop to -273.14 °C.

QF = Quality of the optical image which is individual for every application. Quality factor indicates the

steepness of the slope in the optical image and thus the stability of the measurement. However, in some

applications QF values might be approx. 40-50 which is still good, whereas in other applications the QF is up

to 170. Therefore, it is important to keep track on the QF value so that the changes in QF are recorded.

Generally, significantly decreased QF values are seen with bad optical images that have resulted in prism

coating or erosion.

LED = Optimal value for LED is between 3 and 25 %. Higher LED values indicate blockage on the light path,

which can be a result of coating or prism erosion. High LED values are often shown with bad optical images,

i.e. with weak slopes.

Advertising