empty. Washing when the pipe is empty can by very corrosive to the prism, as there is no counter pressure

in the pipe.

3.1

Wash Recommendations

The other critical part of the wash is the wash medium characteristics, especially the temperature and the

pressure. Below you can find general guidelines for the wash parameters and pressures depending on the

wash medium and nozzle type. It is also recommended that the wash medium temperature should always be

higher than the process medium temperature to avoid any solid formation or crystallization.

Figure 8

Recommended wash pressures and times

3.2

Successful Wash

As the wash is essential part of the fully functional refractometer, regular testing of the prism wash is

recommended. The manual washing is be activated by pressing MENU  3 SENSOR STATUS  MANUAL

WASH. The optical image should change clearly during the wash. Also the temperature reading should