the sensor.

Solution: The cables should be inspected, and if the problem is not resolved, the sensor processor card

needs to be replaced by K-Patents or a trained service person. If the sensor has leaked liquid, most likely all

the electronics and optics needs to be replaced.

4.5 Description: Inaccurate measurement

Cause(s): These issues are generally related to prism coating, which may result in a constantly increasing

concentration reading. It is also common that the fault is found in the reference measurement method (lab

instrument).

Solution: Clean the prism of the sensor carefully and check the optical image. Proceed with the built-in

verification procedure. If the verification fails, the instrument needs to be inspected by qualified person

before recalibration. If the instrument passes, it is an indication that the issue is related to the process

conditions (too low flow velocity) or ineffective washing.

Figure 11

Verification procedure for checking the calibration