Programming, Date, Time – Lencore Spectra Classic: LP-2 User Manual

Page 6

Advertising
background image


-6-

steps.

7.

Locate the

Test/Normal switch on the P.C. Board. Put the switch in Test

position. The display will change showing both a dB level and a voltage level
for each zone. This bypasses any programmed attenuation and allows for a
total system calibration of the LM-6 master units. Adjust the Zone 1

slide

control for a stable reading of 0 dB. Do the same for Zone 2.


8.

The dB levels present here represent the change in decibels from 47dB. You
can either raise or lower the dB level (from this 47dB) here. With the

Test/Normal switch in Test position, adjust all the LM-6 master units for an

office sound level reading of 47 dB's. When all the masters have been

adjusted and tuned as necessary, put the

Test/Normal switch in Normal

position. The display will change to the date and time screen. The dB
readings may change as any programmed attenuation is taken into

account.


9.

Program the timer as per customer specifications. Refer to the programming
section of this manual.



PROGRAMMING


The following keyboard key combinations are documented in a way that assumes
you are starting from the main screen. There will be situations where it is much
quicker to go from one sub-menu to another without returning to and starting from
the main screen. You may use these quicker routes at any time. However, for the
purpose of documentation, the main screen will be the starting point.

Date

Press 0 (for the menu screen)
Press 1

Enter month (two digits)
Enter day (two digits)
Enter year (four digits)
Press 0

Time

Press 0 (for the menu screen)
Press 2
Press 1

Enter hour (two digits)

Enter minute

(two digits)

Enter seconds

(two digits)

Enter 0 for A.M. or 1 for P.M.

Advertising