Section 5

Command Descriptions

5.80

Address change

6

2

G0: Address status change has not occurred.
G1: Address status change has occurred.

Addressed state

9

1

T:

Talker

L:

Listener

I:

Idle

Service Request

11

2

S0: SRQ is not asserted.
S1: SRQ is asserted.

Error code

14

3

Enn:
Letter 'E' followed by two-digit error code.
Refer to Appendix B for Error explanations.

Triggered

18

2

T0: No IEEE Trigger command received.
T1: Received the IEEE Trigger command.

Cleared

21

2

C0: No IEEE Clear command received.
C1: Received the IEEE Clear command.

Error description

24

17

Text of error message

The Operating Mode (C or P) indicates whether or not the Micro488/EX is the

Active Controller. If the Micro488/EX Passes Control to another device, then the
Operating Mode indicator will change from "C" to "P". When the Micro488/EX
regains control, then the indicator will again be "C". If the Micro488/EX is not the
System Controller, then it will initially be a Peripheral and thus the indicator will be
"P". It will, of course, become "C" when the Micro488/EX receives control from the
Active Controller.

The Bus Address is the IEEE bus device address assigned to the Micro488/EX by

the internal hardware switch.

The Address Change (G0, G1) indicator is set whenever the Micro488/EX

transitions from the idle state to a Talker or Listener, or from a Talker or Listener state
to an idle state. It will not indicate when a change is made from a listener to a talker or
a talker to a listener. The address change is reset when STATUS 1 is read.