2 ieee bus terminator selection, Sw2 view for serial terminator selection – Measurement Computing Micro 488/EX rev.2.1 User Manual
Page 23

Section 2
Getting Started
2.9
SW2 View for Serial Terminator Selection
OPEN
1 2 3 4 5 6 7 8
DOT
Switch
Side
View
OPEN
1 2 3 4 5 6 7 8
CR Only
LF-CR
OPEN
1 2 3 4 5 6 7 8
OPEN
1 2 3 4 5 6 7 8
LF Only
CR-LF
2.4.2 IEEE Bus Terminator Selection
SW3-6 through SW3-8 set the IEEE bus terminators used for data sent
or received (Pass-Thru modes only) by the Micro488/EX. EOI, a line used
to signal the end of a multiple character bus transfer, may also be enabled.
If enabled, EOI is asserted when the last selected bus terminator is sent.
Factory default is CR-LF with EOI disabled.
SW3 View for IEEE Bus Terminator Selection
OPEN
1 2 3 4 5 6 7 8
DOT
Switch
Side
View
OPEN
1 2 3 4 5 6 7 8
CR Only
LF-CR
OPEN
1 2 3 4 5 6 7 8
OPEN
1 2 3 4 5 6 7 8
LF Only
CR-LF
OPEN
1 2 3 4 5 6 7 8
OPEN
1 2 3 4 5 6 7 8
EOI Disabled
EOI Enabled