5 Dispensing

with

PD12


For optimal dispensing with the PD12, the following should be observed:

5.1

Vessel Placement ................................................................................................................. 10

5.2

Tube size .............................................................................................................................. 11

5.3

Nature of fill media................................................................................................................ 12

5.4

Priming tubes........................................................................................................................ 12

5.5

Drip....................................................................................................................................... 12

5.6

Hard Feed............................................................................................................................. 12


5.1 Vessel

Placement


In order to build up adequate pressure and reduce friction, it is recommendable to place the
vessel containing fill media at the same level as pump head or preferably above the pump
head level. Placing the vessel higher than pump head level provides positive product support
and may reduce the calibration interval. It is also recommended to place the vessel as close
as possible to the pump on suction side.

Flexicon

Pump Head

Preferred Placement

of container

Normal Placement

of container

Suction side

Pressure side

Fig. 5.1