Instruction Manual

760008-A
September 2001

1-4 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 TO2

1-5 SPECIFICATIONS

1

a. General

Measurement Species................................... Trace Oxygen
Ranges .......................................................... 0 to 100 ppm (output scalable down to 0-2 ppm fullscale)
Accuracy........................................................

±3% of reading or ±0.02% of range (except for ranges ≤

100 ppm:

±3% of reading or ±0.05% of range)

Sensitivity ...................................................... <10 ppb Oxygen
Noise ............................................................. 1% of fullscale, peak to peak
Linearity .........................................................

±1% of fullscale

Response Time ............................................. Typically 90% in less than 20 seconds
Zero Drift........................................................

≤±1% of fullscale/24 hours at constant temperature

Span Drift.......................................................

≤±1% of fullscale/24 hours at constant temperature

Effect of Temperature.................................... 0.32% of reading per

°F from 70°F (0.58% of reading per

°C from 21°C)

Effect of Flow.................................................

≤2% of reading for a flow change of ±250 cc/min (0.5

SCFH)

Operating Temperature ................................. 32

°F to 113°F (0°C to 45°C)

Power Requirements..................................... +24 VDC

±5%, 10 W max.

Ripple and Noise ........................................... <100 mV peak to peak
Line and Load Regulations............................ <

±1%

b. Sample

Sample .......................................................... Non-flammable (below 100% of the LEL)
Flow Rate ...................................................... 0.5 to 1.5 L/min.
Supply Pressure ............................................ 1027 to 1082 hPa - absolute (0.2 to 1.0 psig)
Temperature .................................................. 32

°F to 113°F (0°C to 45°C)

Particulates.................................................... filtered to <0.1 mg/L; non-condensing at ambient tem-

perature

Sample Humidity ........................................... non-condensing at ambient temperatures

c. Physical

Materials in contact with sample ................... Stainless steel, Teflon, Delrin, neoprene
Dimensions.................................................... See Figure 2-2, Outline and Mounting Dimensions
Weight ........................................................... 6.8 kg (15 lbs.)
Mounting........................................................ Horizontal, external to Platform or custom installed in a

panel

Case Classification........................................ General Purpose for installation in weather protected

area

Max. Separation from Platform...................... 1600 m (1 mile)