3. startup & initialization, Instruction manual – Emerson NGA 2000 TO2 User Manual

Page 30

Advertising
background image

Instruction Manual

760008-A
September 2001

3-6 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 TO2

3-3

STARTUP & INITIALIZATION

Establish sample or zero gas flow through the
analyzer module at a nominal flow rate of 2
SCFH (1 l/min). Allow gas to flow for 15 to 30
seconds before applying power. Apply power
to the TO2 Analyzer Module.

After switching on the TO2, the analyzer will
begin its booting procedure which is apparent
on the LCD screen. The first part of the
initialization procedure is a self check of the
software and analyzer components. Various
displays will show the status of the initializa-
tion including revision notes, “Initializing net-
work interface,” “Searching for nodes,”
“Querying Module: TO2, 12% Complete,” and
“Calculating bindings.”

Pressing the F1 key during initializing will re-
set the LCD brightness and contrast to factory
settings (See Section 3-4e). Pressing the F3
key will abort the network initializing, aborting
any connection to other analyzers. In that
case, only the menus of the local analyzer will
be available.

At the end of the initializing routine the
“measure” screen will display as shown on the
next page. This screen is the access to all
other channels, menus and submenus. The
actual display may differ from that shown de-
pending on any custom configuration as de-
scribed in Section 3-5e.

Figure 3-8. Startup Display

LCDReset

(C) 1999 FISHER-ROSEMOUNT Analytical

NGA-2000 Control-Module Rev. 3.6.0 /P010

Language: P012/01/00

Initializing Network

Initializing network interface

Abort