Operation 3-15

Model NGA 2000 TO2

Figure 3-16. Calibration – Gas Scale Factor

Move the cursor to the “Calibration…” line
and press the

↵ key. The “Calibration” menu

will display as shown below.

Move the cursor to the “Gas scale factor…”
line and press the

↵ key. The “Gas Scale

Factor 1/0” menu will display as shown on
the next page.

Enter the volumetric percentages of each
component present in the sample gas.

Move the cursor to the desired line and press
the

↵ key. Then press the ↑ and ↓ keys to

adjust the value and the

← and → keys to

select the digit. Press the

↵ key again to

save the value.

To view and adjust additional background
gas components, press the F5 key to display
the second menu.

After all the background gas concentrations
have been entered, move the cursor to the
“Compute new adjusted gas scale factor!”
command and press the

↵ key. The analyzer

will automatically calculate the new gas scale
factor and display the result.

TO2

37.50 ppm

-- Analyzer Module Setup and Controls --

Calibration…

Alarms…
Ranges…
Linearization…
Measurement display configuration…
Concentration measurement…
Flow measurement…
Temperature measurement…

Measure

Back…

Status…

TO2

37.50 ppm

Measure

Back…

-- Calibration --

Gas scale factor…

Reset calibration!

Calibrate to span gas!

Span gas: 90.0 ppm

Status…

TO2

37.50 ppm

Measure

Back…

-- Gas Scale Factor (1/2) --

Air background gas concentration: 0.0%

C2H4 background gas concentration: 0.0%
CO background gas concentration: 0.0%
CH4 background gas concentration: 0.0%
N2 background gas concentration: 100%
He background gas concentration: 0.0%
H2 background gas concentration: 0.0%
NH3 background gas concentration: 0.0%
C2H6 background gas concentration: 0.0%
C3H6 background gas concentration: 0.0%

TO2

37.50 ppm

-- Gas Scale Factor (2/2) --

C4H10 background gas concentration: 0.0%

C6H14 background gas concentration: 0.0%

Other background gas concentration: 0.0%
Other background gas factor: 100.0%

Compute new adjusted gas scale factor!

Adjusted gas scale factor: 1.00000
Scale factor computation result: OK

Measure

Back…

Main Menu

Analyzer and I/O, expert controls and setup…

Analyzer module controls… (Or Analyzer module setup…)