Instruction manual – Emerson NGA 2000 TO2 User Manual

Page 65

Advertising
background image

Instruction Manual

760008-A

September 2001

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-5

Model NGA 2000 TO2

Select Linearization…

Select Linearization raw values (sensor cur-
rent)…

Enter the data for each raw value point.

Return to the previous menu (Back…)

Select Linearization set point values (ppm
values)…

Enter the data for each set point value.

Return to the previous menu (Back…)

Select Linearization temperature values…

Enter the data for each temperature value
point..

Return the the previous menu (Back…)

Select Load linearization data! press enter.
The new data is now loaded into the analyzer
and accurate measurements can be obtained.

TO2

37.50 ppm

-- Analyzer Module Setup and Controls --

Calibration…
Alarms…
Ranges…

Linearization…

Measurement display configuration…
Concentration measurement…
Flow measurement…
Temperature measurement…

Measure

Back…

Status…

TO2

37.50 ppm

-- View & Modify Linearization Table --

Linearization raw values (sensor current)…

Linearization set point values (ppm values)…
Linearization temperature values…

Load linearization data!

Measure

Back…

TO2

37.50 ppm

-- Linearization Raw Values --

X1: 0.3828 uA

X2: 3.0400 uA
X3: 20.9300 uA
X4: 193.000 uA

Measure

Back…

TO2

37.50 ppm

-- View & Modify Linearization Table --

Linearization raw values (sensor current)…
Linearization set point values (ppm values)…

Linearization temperature values…

Load linearization data!

Measure

Back…

TO2

37.50 ppm

-- Linearization Set Point Values --

Y1: 0.0000 ppm

Y2: 1.050 ppm
Y3: 7.000 ppm
Y4: 72.800 ppm

Measure

Back…

TO2

37.50 ppm

-- View & Modify Linearization Table --

Linearization raw values (sensor current)…

Linearization set point values (ppm values)…

Linearization temperature values…

Load linearization data!

Measure

Back…

TO2

37.50 ppm

-- Linearization Temperature Values --

T1: 25.0 °C

T2: 25.0 °C

T3: 25.0 °C

T4: 25.0 °C

Measure

Back…

TO2

37.50 ppm

-- View & Modify Linearization Table --

Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…

Load linearization data!

Measure

Back…