-26-

After replacing the Power Board, PF.C. Board, readjust
the Output voltage adjustment as follows.

1. Connect a digital voltmeter to pins 1 (+) and 3 (-) of

K6D

.

2. Adjust the voltage by using

VR611

as following.

AC Input

Reading

230V 370V

±2V

or 120V

340V

±2V

Caution:
Be sure to connect the lamp when taking this adjust-
ment.

Circuit Adjustments

CAUTION: The each circuit has been made by the fine adjustment at factory. Do not attempt to adjust the follow-

ing adjustments except requiring the readjustments in servicing otherwise it may cause loss of per-
formance and product safety.

[Adjustment Condition]

Input signal

Video signal .......................... 1.0Vp-p/75

terminated, 16 steps gray scale (Composite video signal)

Computer signal...................... 0.7Vp-p/75

terminated, 16 steps gray scale pattern (XGA)

Picture control mode ...................... “STANDARD” mode unless otherwise noted.

Note:
* Please refer to “Service Adjustment Menu Operation” for entering to the service mode and adjusting the service

data.

Electrical Adjustments

Output Voltage adjustment

1. Receive the 16-step grey scale video signal.
2. Set to VIDEO mode.
3. Enter the service mode.

[R-PEDESTAL ADJUSTMENT]
4. Connect an oscilloscope to test point “

TP201R

” (+)

and chassis ground (-).

5. Select group no. “

3

”, item no. “

14

” and change data

value to adjust the pedestal level and black level to be
the same level.

[G-PEDESTAL ADJUSTMENT]
6. Connect an oscilloscope to test point “

TP201G

” (+)

and chassis ground (-).

7. Select group no. “

3

”, item no. “

15

” and change data

value to adjust the pedestal level and black level to be
the same level.

[B-PEDESTAL ADJUSTMENT]
8. Connect an oscilloscope to test point “

TP201B

” (+)

and chassis ground (-).

9. Select group no. “

3

”, item no. “

16

” and change data

value to adjust the pedestal level and black level to be
the same level.

Pedestal level = Black level

(a)

Pedestal adjustment

1. Enter the service mode and select group no. “

11

” and

item no. “

0

”. Set data value to “

1

”.

2. Connect a digital voltmeter to test point “TP12V1” (+)

and chassis ground (-). Select item no. “

1

” and change

data value to adjust voltage to be 8.0 ±0.1V.

3. Connect a digital voltmeter to test point “TP12V2” (+)

and chassis ground (-). Select item no. “

2

” and

change data value to adjust voltage to be 9.0 ±0.1V.

4. Select group no. “

11

” and item no. “

0

” and set data

value to “

3

”.

5. Connect a digital voltmeter to test point “TP12V1” (+)

and chassis ground (-). Select item no. “

3

” and change

data value to adjust voltage to be 13.0 ±0.1V.

6. Connect a digital voltmeter to test point “TP12V2” (+)

and chassis ground (-). Select item no. “

4

” and change

data value to adjust voltage to be 13.5 ±0.1V.

Fan Voltage adjustment