Eiki LC-NB3E User Manual

Page 28

Advertising
background image

-28-

Electrical Adjustments

1. Receive the 16-step gray scale computer signal.
2. Set to COMPUTER mode.
3. Enter the service mode.

[R-GAIN ADJUSTMENT]
4. Connect an oscilloscope to test point “

TP25R1

” (+)

and chassis ground (-).

5. Select group no. “

2

”, item no. “

3

” and change data

value to adjust waveform “a” to be minimum ampli-
tude.

[G-GAIN ADJUSTMENT]
6. Connect an oscilloscope to test point “

TP25G1

” (+)

and chassis ground (-).

7. Select group no. “

2

”, item no. “

4

” and change data

value to adjust waveform “a” to be minimum ampli-
tude.

[B-GAIN ADJUSTMENT]
8. Connect an oscilloscope to test point “

TP25B1

” (+)

and chassis ground (-).

9. Select group no. “

2

”, item no. “

5

” and change data

value to adjust waveform “a” to be minimum ampli-
tude.

White Level

(a)

PC Gain adjustment

1. Receive the 16-step gray scale computer signal.
2. Set to COMPUTER mode.
3. Enter the service mode.

[R-OFFSET ADJUSTMENT]
4. Connect an oscilloscope to test point “

TP25R1

” (+)

and chassis ground (-).

5. Select group no. “

3

”, item no. “

11

” and change data

value to adjust the waveform “a” (black portion ) to
be maximum amplitude.

[G-OFFSET ADJUSTMENT]
6. Connect an oscilloscope to test point “

TP25G1

” (+)

and chassis ground (-).

7. Select group no. “

3

”, item no. “

12

” and change data

value to adjust the waveform “a” (black portion ) to
be maximum amplitude.

[B-OFFSET ADJUSTMENT]
8. Connect an oscilloscope to test point “

TP25B1

” (+)

and chassis ground (-).

9. Select group no. “

3

”, item no. “

13

” and change data

value to adjust the waveform “a” (black portion ) to
be maximum amplitude.

Black Level

(a)

PC Offset adjustment