RIGOL DP832 User Manual

Page 61

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DP800 User’s Guide

2-11

You can perform the following operations according to your need.

1. Only enable the track function of a single channel and track the

voltage setting value

The operation procedures are as follows (in this example, only the track function

of CH1 of DP832 is enabled).
 Select the "Independent" tracking mode

Press UtilitySystemTrack SetTrack to select "Independent".

 Enable the track function of CH1

Select CH1 and press Track to select "On". At this point, the track function

of CH1 is enabled and the tracking status icon

is displayed between

the CH1 and CH2 areas in the user interface.

 Disable the track function of CH2 (ignore this step if the track function of

CH2 is currently disabled)

Select CH2 and press Track to select "Off". At this point, the track function

of CH2 is disabled.


 Track the voltage setting value

Select CH1, press Voltage and set the desired voltage. At this point, the

voltage of CH2 changes accordingly. For example, set the voltage of CH1 to

+5V; the voltage of CH2 will change to +5V automatically.

Note: At this point, the voltage setting value of CH2 cannot be set and can

only change with that of CH1.

2. Enable the track functions of both of the two channels and track the

voltage setting value

First, you can enable the track functions of the two channels using any of the

following methods (take DP832 as an example).

Method 1
 Select the "Synchronous" tracking mode

Press UtilitySystemTrack SetTrack to select "Synchronous".

 Enable the track functions of CH1 and CH2 at the same time

Select CH1 or CH2, press Track and select "On". At this point, the track

functions of CH1 and CH2 are both enabled. The tracking status icon

is

Tip

The track function only tracks the voltage setting value and the actual output

voltage will not be affected.

Advertising