Sense working mode. The Sense connecting method at the rear panel is as shown in

the figure below.

Operation Procedures:

1. Connect the load to the corresponding position on the connecting terminal

correctly as shown in the figure above. Pay attention to the polarity when

making connections.

2. Insert the connecting terminal into the output interface at the rear panel of the

instrument correctly. Pay attention to the corresponding relations of the

connecting terminal and the output interface.

3. After selecting the desired range, press Sense to turn on the Sense function.

Note: The output terminals at the front panel and the output interface at the rear

panel cannot be used for the Sense mode at the same time. You can only select one

of them for the Sense function at each time (wherein, the output terminals at the

front panel provide higher output accuracy).