Amprobe AM-220 Digital-Multimeters User Manual

Page 14

Advertising
background image

10

BATTERY REPLACEMENT
1) When the battery voltage drops below proper operating range, the “

” symbol

will appear on the LCD and the batteries needs to be changed.

2) Before changing the batteries, set the selector switch to “OFF” position. Remove the

two screws on the bottom case and lift the bottom case.

3) Replace the old batteries with the same type battery.
4) Close the bottom case and fasten the screw.

FUSE REPLACEMENT
1) This meter is provided with a 0.5A/250V fuse to protect the current measuring

circuits that measure up to 400mA, with a 10A/250V fuse to protect the 10A range.

2) Ensure the instrument is not connected to any external circuit, set the selector switch

to “OFF” position and remove the test leads from the terminals.

3) Remove the two screws on the bottom case and lift the bottom case. Replace the old

fuse with the same type and rating: 5x20mm 0.5A/250V or 6x32mm 10A/250V fuse.

4 Close the bottom case and fasten the screws.

MAINTENANCE
1) Before opening the bottom case, disconnect both test leads and never use the meter

before the bottom case is closed.

2) To avoid contamination or static damage, do not touch the circuit board without

proper static protection.

3) If the meter is not going to be used for a long time, take out the batteries and do

not store the meter in high temperature or high humidity environment.

4) Repairs or servicing not covered in this manual should only by qualified personnel.
5) Periodically wipe the case with a damp cloth and detergent. Do not use abrasives or

solvents on the meter.