Amprobe TMA-21HW Hot-Wire-Anemometer User Manual

Page 18

Advertising
background image

6

• Push “” and “” keys to scroll through the log interval choices. The

choice are : 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25 and 30 seconds, and 1, 2,
3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 minutes. Push “↵” key to store
the choice.

• Push “SET” key several times until the “ SET ” symbol is disappeared to

exit the setting mode.

Auto Power Off Time Setting mode

• Push “SET” key four times to enter this mode, the “APO” symbol and the

current auto power off time are displayed.

• Push “” and “” keys to setting the desired auto power off time from

1 to 50 minutes or setting to “– – m” for disable this function. Push “↵”
key to store the setting.

• Push “SET” key several times until the “ SET ” symbol is disappeared to

exit the setting mode.

Backlighting Time Setting mode

• Push “SET” key five times to enter this mode, the “bL” symbol and the

current backlight time are displayed.

• Push “” and “” keys to setting the desired backlight time from 1 to

50 seconds or setting to “– – s” for disable this function. Push “↵” key to
store the setting.

• Push “SET” key several times until the “ SET ” symbol is disappeared to

exit the setting mode.

Barometric Pressure Setting mode

• Push “SET” key eight times to enter this mode, the “ SET ” and “mmHg

or “inHg” symbols are displayed.

• Push , ,  and  keys to setting the actual barometric pressure value.

• Push “↵” key to store the value.

• Push “SET” key several times until the “ SET ” symbol is disappeared to

exit the setting mode.

The actual barometric pressure must be entered to convert air velocity

and volume measurements to actual conditions.

Advertising