Specifications – Amprobe GP-1 Earth-Tester-Ground-Probe User Manual

Page 18

Advertising
background image

RESISTANCE
Range

Resolution

0 - 19.99

0.01

20.0 - 199.9

0.1

200 - 1999

1

Accuracy:

+/- (2% of reading + 2 Digits)

RESISTIVITY
Range

Resolution

0-19.99

m

0,01

20.0-199.9

m

0.1

200-1999

m

1

2.00k - 19.9k

m

10

20.0k - 199.9k

m

100

200k - 314

m

1000

Accuracy:

+/- (2% of reading + 2

π

a.0.02

); where

ρ ≤

19.99

2

π

a

+/- (2% of reading + 2

π

a.0.2

); where 19.99

<

ρ ≤

199.9

2

π

a

+/- (2% of reading + 2

π

a.2

); where 19.99

<

ρ

2

π

a

Test current:

<10mARMS sine wave

Test Frequency:

135 Hz+/- 1Hz

VOLTAGE
Range:
0-600VRMS max

+/- (2% of reading + 2 Digits)

INCLUDED ACCESSORIES
Model GP-CS”

(1) 30’(10m) Green test cable
(1) 30’(10m) Black test cable
(1) 47’(15.5m) Red Test cable
(1) 77’(25.5m) Blue test cable

Model GP-MR

(4) 9”measuring rods

Model GP-TL

(1) Safety banana test lead set

Model GP-RS232

(1) 9 pin male to 9 pin female-RS-232 cable

Model GP-BP

(1) Black shoulder pouch

Download Software

GENERAL
Fuse:

5 x 20mm, 32mA @250V fast blow, Amprobe Model 5 x 20-250.32

Input Power:

(4) “D” cells, Amprobe Model MN-1300

Display:

3-1/2 Digit L.C.D. 3/4”(H)

Weight:

7.26 lbs. (3.3kg)

Dimensions(WxHxL):

13.6”x5.1”x9.8” (3.45Cm x 13Cm x 25cm)

Nominal Temp. Range:

41 - 95°F (5 - 35°C)

Working Temp. Range:

32 - 104°F (0 - 40°C)

Storage Temp. Range:

14 - 140°F (-10 - 60°C)

STANDARDS
Constructed according to:

EN61010-1

Ground resistance measurement:

VDE 0413 Part 7

Radio frequency emission

EN50081-1

Radio frequency immunity

EN50082-1 Class B

Specifications

Page 18

Advertising