Automatic Operation (CA-3000, CA-3200 models only):

1.

Attach the appropriate interface to the patient.

2.

To operate the unit automatically, set the manual/auto switch to the auto position.

The unit will cycle from inhale (positive) to exhale (negative) to zero pressure, and

back to inhale.

3.

After 4 to 5 cycles, set the manual/auto switch back to the manual position. Remove

the patient interface from the patient and allow time for a normal breathing pattern to

return (20 to 30 seconds), or place the patient back on the ventilator if currently in

use. Avoid prolonged periods connected to the device. During this resting period, clear

secretions that may have become visible in the mouth, throat or tracheostomy tube.

CAUTION: This unit is designed for Intermittent Operation Only and not for continuous

use. The device should not be cycled continuously for more than 5 minutes. After such

time, the unit should either be turned off or left idling with the blower on for at least

5 minutes.

O

P E R AT I O N

V

E R I F I C AT I O N

(

A L L M O D E L S

)

It is recommended that the CoughAssist MI-E be periodically tested to ensure that the

cycling valve returns to the neutral, or pause, position after either the inhale or exhale

phase. To determine this, follow these steps:

1.

Attach a patient circuit to the unit and block the end of the hose.

2.

Turn the power switch ON.

3.

Set the manual/auto switch to the manual position (automatic models only).

4.

Set the pressure knob fully clockwise (maximum pressure).

5.

Cycle the manual control lever from inhale to exhale and observe the pressure gauge

to ensure that positive and negative pressure is being applied to the patient circuit.

6.

Release the manual control lever from the inhale position and observe that the pressure

immediately drops to 0 cm H

2

O. Repeat for the exhale position. In either case, if the

pressure does not drop to zero, the unit should be returned for repair.

13