Class 1 Equipment: Equipment in which protection against electric shock does not

rely on basic insulation only, but includes a grounding pin on the power cord. For

ground reliability always plug the power cord into an AC grounded outlet.

Type BF Equipment: Type B piece of equipment with an F-Type applied part.

A Type B piece of equipment is one that provides a particular degree of protection

against electric shock, particularly regarding allowable leakage current and reliability

of the protective earth connection (grounding).

F-Type applied part is one that extends from the patient into the equipment and is

isolated from all other parts of the equipment.

Water Ingress: This device does not have protection against ingress of water.

Disinfection: With the exception of the patient circuit, this device can be disinfec-

ted using 70% isopropyl alcohol or equivalent. (See Section 6: Cleaning and

Disinfection.)

Flammable Anesthetics: This device is not suitable for use in the presence of a

flammable anesthetic mixture with air, or in the presence of a flammable anesthetic

mixture with oxygen or nitrous oxide.

Intermittent Operation: This device is designed for Intermittent Operation Only

and not for continuous use. The device should not be cycled continuously for more

than 5 minutes. After such time, the unit should either be turned off or left idling with

the blower on for at least 5 minutes.

E Q U I P M E N T C L A S S I F I C AT I O N

Per IEC 60601-1, Medical Electrical Equipment, General Requirements for Safety, the
CoughAssist MI-E is classified as follows:

SECTION 2

W

A R N I N G S

A N D

C

A U T I O N S

3

92804 3/28/01 8:35 AM Page 5