1. Manual/Auto: Changes the cycling mechanism to either manual or auto-

matic mode.

2. Inhale: Sets time interval (sec) for Inhale phase of automatic cycling.

Not operative in the manual mode.

3. Exhale: Sets time interval (sec) for Exhale phase of automatic cycling.

Not operative in the manual mode.

4. Pause: Sets time interval (sec) for Pause phase of automatic cycling.

Not operative in the manual mode.

5.

Power: The power switch turns on or off the CoughAssist MI-E. The “I” symbol on the

switch designates the on position. When activated a green light within the switch will

illuminate. The “O” symbol designates the off position.

6.

Manual Control Lever: Use to manually cycle the unit to inhale or exhale. Not operative

in the automatic mode.

7.

Pressure: Varies the inhalation and exhalation pressures together (also see Inhale Pressure).

8.

Handle: Recessed carrying handle.

9.

Patient Port: Connection for patient circuit.

10.

: Symbol for Type B Equipment with F-Type applied part.

11. Inhale Flow: There are two inhalation flow settings: Full ( ) and Reduced ( ).

NOTE: When using reduced inhalation flow there will be a small reduction in inhale pressure.

12. Inhale Pressure: Varies the inhalation pressure between 50% and 100% of the

exhale pressure.

13. Pressure Gauge Zero Adjust: Access to “zero” adjustment on the pressure gauge.

Use only if the gauge does not return to “O” when the unit is turned off (see Section 8).

14. Pressure Gauge: Indicates pressure in the patient circuit calibrated in cm H

2

O.

SECTION 4

7

C

O N T R O L S

,

C O N N E C T O R S

&

V I S U A L I N D I C AT O R S

o