T

R O U B L E S H O O T I N G

G

U I D E

1.

Pressure Gauge Adjustment: If the pressure

gauge does not go to “0” when the device is turned

off it must be readjusted. Remove the adhesive cover

over the pressure gauge zero adjust and turn the adjust

screw to “0” using a screwdriver. If unable to make this

adjustment, the unit should be returned for servicing.

2.

Fuse Replacement: If the unit is connected to the

proper power source and the green light within the

power switch does not illuminate when the switch is

actuated, one or both of the two safety fuses may have

blown. The procedure for replacement of a blown fuse

is as follows:

a.

Disconnect the unit from any power outlet and

disconnect the power cord from the receptacle on

the rear of the unit (see Back Panel diagram on

page 9).

b.

Locate the access door on the receptacle labeled

with the symbol:

. Open the access door

by prying the latch at the top with a small screw

driver or fingernail. Pivot the door down to reveal

the two fuse holders.

c.

Press each of the two spring clips to the side

(the left one to the left, the right one to the right)

and slide both fuse holders out of the receptacle.

d. Inspect both fuses and replace, if necessary, with

fuses with equivalent ratings, as shown on the

rear Product Label Description (page 9). To

replace a fuse, slide the damaged one out of the

holder and slide the new fuse in its place.

e. Reinsert each fuse and holder into the receptacle,

close the access door and reconnect the power cord.

17

Door Latch

Spring Clip

Fuse Holder

Access Door

Spring Clip

Fuse Holder