1.

Pressure: Varies the inhalation and exhalation pressures together

(also see Inhale Pressure).

2.

Manual Control Lever: Use to manually cycle the unit to inhale

or exhale.

3.

Power: The power switch turns on or off the unit. The “I” symbol

on the switch designates the on position. When activated a green

light within the switch will illuminate. The “O” symbol designates

the off position.

4.

Handle: Recessed carrying handle.

5.

Patient Port: Connection for patient circuit.

6.

: Symbol for Type B Equipment with F-Type applied part.

7.

Inhale Flow: There are two inhalation flow settings: Full ( ) and Reduced ( ).

NOTE: When using reduced inspiratory flow there will be a small reduction in inspiratory pressure.

8.

Inhale Pressure: Varies the inhalation pressure between 50% and 100% of the

exhale pressure.

9.

Pressure Gauge Zero Adjust: Access to “zero” adjustment on the pressure gauge.

Use only if the gauge does not return to “0” when the unit is turned off (see Section 8).

10. Pressure Gauge: Indicates pressure in the patient circuit calibrated in cm H

2

O.

o

1

2

3

4

5

6

7

8

9

10