Storing, Logging and Printing

Logging Results

Logging Results

Description

Test Period Logging

If degradations in system performance can be observed at an early stage,
then the appropriate remedial action can be taken to maximize circuit
availability and avoid system crashes. Test period logging allows you to
monitor the error performance of your circuit. At the end of the test
period the selected results are logged. Results can be logged at regular
intervals during the test period by selecting a LOGGING PERIOD of
shorter duration than the test period. An instant summary of the results
can be demanded by pressing

without affecting the test in

progress.

Error Event Logging

Manual tracing of intermittent faults is time consuming. Error event
logging allows you to carry out unattended long term monitoring of the
circuit. Each occurrence of the selected error event is logged.

The results obtained during the test are retained in memory until they
are overwritten by the next set of results. The results can be logged at
any time during the test period and at the end of the test period. The
results required are selected using

LOGGING

SETUP

.

Any Alarm occurrence results in a timed and dated message being
logged.

BER and Analysis results can be selected by the user.

Cumulative and Period versions of the results are calculated and can be
selected by the user.

Period

The results obtained over a set period of time during
the test. The Period is defined by the LOGGING
PERIOD selection.

Cumulative

The results obtained over the time elapsed since the
start of the test.

The results can be logged to the following devices, selectable using

LOGGING SETUP

:

PRINT NOW

OTHER

LOGGING

CONTROL

OTHER

LOGGING

DEVICE