When “SET” is pressed the display shows the set-point memory that is cur-
rently being used and its value. Eight set-point memories are available.

S

Access set-point selection

1. 25.0

Set-point 1, 25.0°C currently used

Press “SET” to select this memory and access the set-point value.

S

Access set-point value

C 25.00

Current value of set-point 1, 25.00°C

Press “UP” or “DOWN” to change the set-point value.

U

Increment display

C 30.00

New set-point value

Press SET to accept the new value. The bath begins heating or cooling to the
new set-point.

S

Store new set-point

Press “EXIT” and the bath temperature will be displayed again.

E

Return to the temperature display

24.73 C

Bath temperature display

The bath heats or cools until it reaches the new set-point temperature. Turn off
the cooling to reach and control at higher temperatures.

When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid.

To obtain optimum control stability adjust the proportional band as discussed in
Section 9.6.

15

4 Quick Start