4 linefeed 45, 11 calibration parameters 45, 4 linefeed – GE CTR 80 User Manual

Page 48: 11 calibration parameters

Advertising
background image

dUPL

Flashes “dUPL” and then displays the setting

d=FULL

Current duplex mode setting

The mode may be changed using “UP” or “DOWN” and pressing “SET”.

d=HALF

New duplex mode setting

9.10.4

Linefeed

The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,

LF

Flashes “LF” and then displays the setting

LF= On

Current linefeed setting

The mode may be changed using “UP” or “DOWN” and pressing “SET”.

LF= OFF

New linefeed setting

9.11

Calibration Parameters

The operator of the CTR –80 controller has access to the Hard Cutout and a
number of the bath calibration constants namely R

0

, ALPHA, DELTA, and

BETA. These values are set at the factory and should not be altered. The cor-
rect values are important to the accuracy and proper and safe operation of the
bath. Access to these parameters is available to the user only so that in the
event that the controller memory fails the user may restore these values to the
factory settings. The user should have a list of these constants and their settings
with the manual.

Caution: DO NOT change the values of the bath calibration constants

from the factory set values. The correct setting of these parameters is im-
portant to the safety and proper operation of the bath.

The calibration parameters menu is indicated by:

CAL

Calibration parameters menu

Press “SET” five times to enter the menu.

45

9 Controller Operation