7 cutout 41, 7 cutout – GE CTR 80 User Manual

Page 44

Advertising
background image

9.7

Cutout

As a protection against software or hardware fault or user error, the bath is
equipped with an adjustable cutout device that shuts off power to the heater if
the temperature exceeds a set value. This protects the instrument, probes, and
fluid from excessive temperatures. The cutout temperature is programmable by
the operator from the front panel of the controller. The cutout should be set to
the temperature limits of the selected fluid.

If the cutout is activated because of excessive temperature, power to the heater
shuts off and the instrument cools. The bath cools until it reaches a few degrees
below the cutout set-point temperature. At this point the cutout resets and al-
lows normal operation.

The cutout set-point may be accessed within the secondary menu. Press “SET”
and “EXIT” to enter the secondary menu and show the heater power. Then
press “SET” twice to access the cutout set-point.

S+E

Access heater power in secondary menu

12.0 P

Heater power in percent

S

Access proportional band

Pb= 0.05C

Proportional band setting

S

Access cutout set-point

CO= 80C

Cutout set-point

To change the cutout set-point press “UP” or “DOWN”.

CO= 70C

New cutout set-point

To accept the new cutout set-point press “SET”.

S

Accept cutout set-point

The next function is the configuration menu. Press “EXIT” to resume display-
ing the temperature.

41

9 Controller Operation

Advertising