Table 3 controller communications commands 50, Table 3, Controller communications commands – GE CTR 80 User Manual

Page 53

Advertising
background image

50

10 Digital Communication Interface

Command Description

Command
Format

Command
Example

Returned

Returned
Example

Acceptable
Values

Display Temperature

Read current set-point

s[etpoint]

s

set: 999.99 {C or F}

set: 90.00 C

Set current set-point to n

s[etpoint]=n

s=80.00

Instrument
Range

Read temperature

t[emperature]

t

t: 999.99 {C or F}

t: 55.6 C

Read temperature units

u[nits]

u

u: x

u: C

C or F

Set temperature units to Celsius

u[nits]=c

u=c

Set temperature units to
Fahrenheit

u[nits]=f

u=f

Read scan mode

sc[an]

sc

scan: {ON or OFF}

scan:ON

Set scan mode

sc[an]=on/off

sc=on

ON or OFF

Read scan rate

sr[ate]

sr

srat: 99.9 {C or F}/min

srat:12.4C/min

Set scan rate

sr[ate]=n

sr=1.1

.1 to 99.9

Secondary Menu

Read proportional band setting

pr[opband]

pr

pb: 999.9

pb: 15.9

Set proportional band to

n

pr[opband]=n

pr=8.83

Depends on
Configuration

Read cut-out setting

cu[tout]

cu

cu:9999 {C or F}

cu:105 C, in

Set cut-out setting to

n degrees

cu[tout]=

n

cu-95

25 to 115°C

Read heater power

(duty cycle)

po[wer]

po

po: 999.9

po: 1.0

Configuration Menu

Read high limit

hl

hl

hl:999

hl:126

Set high limit

hl=n

hl=90

25 to 100

Read low limit

ll

ll

ll:999

ll:-90

Set low limit

ll=n

ll=-90

–90 to 25

Read cooling setting

co[ol]

co

cool: {ON or OFF}

cool: ON

Set cooling setting

co[ol]=on/of[f]

co=on

ON or OFF

Read serial sample setting

sa[mple]

sa

sa: 9

sa: 1

Set serial sampling setting to

n

seconds

sa[mple]=n

sa=0

0 to 999

FULL or HALF

Set serial duplex mode to full

du[plex]=f[ull]

du=f

Set serial duplex mode to half

du[plex]=h[alf]

du=h

Table 3

Controller Communications Commands