3 cooling 43, 10 serial interface parameters 43, 3 cooling – GE CTR 80 User Manual

Page 46: 10 serial interface parameters

Advertising
background image

LL

Low Limit parameter

Press “SET” to enable adjustment of LL.

LL

Flashes “

” and then displays the setting

L=-90

Current LL setting

Adjust the LL parameter using “UP” or “DOWN”.

L=-20

New LL setting

Press “SET” to accept the new temperature limit.

9.9.3

Cooling

This menu function allows the operator to disable cooling and corresponds to
the remote cooling control function (see Table 3). Switching the cooling off
temporarily allows the bath to heat up more quickly from a low temperature.
Cooling can be manually switched off using the cooling switch on the front
panel or automatically switched off any time the bath temperature or set-point
is higher than about 35°C. As a result, cooling is ON only if the cooling func-
tion is ON, the cooling switch is ON, and both the bath temperature and the
set-point are less than about 35°C. The cooling function is indicated by,

CooL

Flashes “CooL” and then displays the setting

CooL=OFF

Current cooling setting

The setting may be changed using “UP” or “DOWN” and pressing “SET”.

CooL=On

New cooling setting

9.10

Serial Interface Parameters

The serial RS-232 interface parameters menu is indicated by,

SErIAL

Serial RS-232 interface parameters menu

The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to instruments fitted
with the serial interface. The parameters in the menu are — BAUD rate, sample
period, duplex mode, and linefeed. Press “UP” to enter the menu.

43

9 Controller Operation

Advertising