1)

The system fuses are 20 amp SB, 250V for 115 VAC operation and 10
amp ST, 250 V for 230 VAC operation.

2)

The power cord is rated at 115 VAC, 20 amps. (230 VAC, 10 amps op-
tional.)

3)

A drain plug is provided for ease of removing the fluid media from the
bath. Always use a container of adequate size to hold the FULL LOAD
of fluid. Some oils are more easily drained at higher temperatures.

Caution: Do not exceed a 100°C fluid temperature for draining. The valve

could be damaged if 100°C is exceeded. Insulate the container from the
floor and other objects.

4)

The serial RS-232 interface attaches to the back of the bath at the con-
nector labeled "RS-232".

5)

The removable vent panel can be removed to access the condensor for
cleaning. See Section 12, Maintenance.