6. calibration certification procedure – Miller Electric IH User Manual

Page 69

Advertising
background image

OM-203 185 Page 65

8-6. Calibration Certification Procedure

The following items are required for verifying calibration:

1. Fluke-714 Thermocouple Calibrator

this can be obtained from: http://www.fluke.com/products/home.asp?SID=13&AGID=9&PID=1707

specify primary standard certification when ordering.

2. Type-K Mini Connector (Fluke #80CK-M) Or Equivalent

this can be obtained from: http://thermo-electric-direct.com.control.interliant.com/

Miniature_Quick-Coupling_Thermocouple_Connectors10630.html

3. Type-K 2-Pin Male Connector

this can be obtained from: http://thermo-electric-direct.com.control.interliant.com/

Standard_Quick-Coupling_Thermocouple_Connectors7886.html

4. Calibration Labels − Part No. QCC306BU BLUE

these can be obtained from: http://www.q-cees.com/itemdetail.asp?part=206

5. A length of type-K thermocouple wire connected between the type-K mini connector and type-K 2-pin

connector.

1. Fluke-714 Thermocouple

Calibrator

2. Type-K Mini Connector
3. Type-K 2-Pin Male Connector
4. Calibration Labels

1

2

3

4

Perform the procedure as follows:

1.

Press the down arrow on bottom left of digital recorder to display numeric page.

.

Allow readings to stabilize before recording.

2.

Set Fluke-714 TC Calibrator to 382.0 degrees F.

a. Move TC Calibrator from TC1 through TC6 in sequence recording each value read onto the Certificate of

Calibration.

b. Verify readings are within

±

3 degrees F of TC Calibrator.

.

If readings are beyond the +3degrees F, the instrument will need to be returned to Eurotherm for repair.

Phone: 703-669-1316

E-mail: [email protected]

web page: www.eurotherm3.com/repairs.htm

c. Repeat at 882.0 degrees F and at 1382.0 degrees F

3.

Save a copy of the Certificate using the IH/TS serial no. and date as the file name, i.e. “LB056241 062501” for

June 25, 2001.

4.

Print a copy of the certificate to send with the IH/TS.

5.

Complete calibration sticker and apply to IH/TS front panel below recorder.

I.D. No. (IH/TS serial no.)

By (your initials) Date (today’s date)

Due (date 1 year from today)

Advertising
This manual is related to the following products: