PROCEDURE 6

ELECTRONICS

REPOSITIONING BATTERY
CHARGER CONNECTOR (FIGURE 3)

NOTE: This procedure applies to wheelchairs equipped
with the Mark IV controller ONLY.

1. Remove the hex bolt from the T-nut that secures the

battery charger connector bracket to the wheelchair.

2. Remove the battery charger connector bracket/

battery charger connector.

3. Slide the T-nut out from the channel on the base

frame of the wheelchair.

4. Reposition the T-nut in the channel on the oppo-

site side of the wheelchair.

5. Install the hex bolt through the battery charger con-

nector bracket AND into the T-nut. Securely tighten.

FIGURE 3 - REPOSITIONING BATTERY CHARGER

CONNECTOR

FIGURE 4 - FUSE REPLACEMENT

Fuse

Fuse

Holder

Battery Charger Connector

Base Frame

Hex Bolt

FUSE REPLACEMENT (FIGURE 4)

NOTE: This procedure applies to wheelchairs equipped
with the Mark IV controller ONLY.

1. Remove the front shroud. Refer to REMOVING/IN-

STALLING THE SHROUDS in PROCEDURE 9 of
this manual.

2. Remove the front battery. Refer to REMOVING/IN-

STALLING THE BATTERIES in PROCEDURE 11
of this manual.

3. Grasp and pull the existing fuse to remove it from

fuse holder.

4. Install new fuse.

5. Reassemble wheelchair by reversing STEPS 1-2.

Channel

Battery Charger

Connector Bracket

T-Nut

NOTE: Base frame shown is for reference ONLY. It is
NOT a true representation of the Pronto base frame.