SIGMA BM-286A User Manual

Page 19

Advertising
background image

3. OPERATING INSTRUCTIONS page 16

Model BM-286A

Minimum

Weight

Lbs:

The minimum weight in pounds required before a scan may be initiated. Enter a

value between 5 and 500 pounds.

NOTE: This value must be input in pounds regardless of the units of weight

selected.

BM

I.D.:

This is the identification for the monitor. This ID will be included in the reports

and data sample information. Enter any combination of eight alpha-numeric

characters.

User

Stream:

This entry allows the system administrator to allow all operators access to any

one, or all 20 of the available waste streams.

If this entry is set to "0", the operator must select the appropriate waste stream for

each item to be scanned.

If this entry is set to 1 - 19, the operator will only be able to use the waste stream

selected by the system administrator.

NOTE: This is a global entry that applies to all operators.