SIGMA BM-286A User Manual

Page 23

Advertising
background image

3. OPERATING INSTRUCTIONS page 20

Model BM-286A

Efficiency:

This parameter entry represents the efficiency of the detectors, in percent, to the

source activity. This value is used in calculating the activity.

This value may be either entered manually or automatically calculated by the

BM-286A. Press on the white box and enter a value, or press "auto" to allow the

unit to calculate the efficiency.

1. Manual entry method:

Enter an efficiency in percent. Entering a <0> will disable the current waste I.D.

2. Automatic calculation:

The operator will be prompted for a count time (in seconds) and the calibration

source activity (in nanocuries).

Press "BEGIN" at which time a START button will appear.

Press the "START" button and the unit will then take an nn second background.

The operator will be prompted to place the source in the center of the chamber

and press "START" to accumulate an nn second count with the source in the

chamber.

The efficiency will be displayed in the efficiency window.