3. OPERATING INSTRUCTIONS page 18

Model BM-286A

The actual formula used to determine minimum scan time is derived from the

RDA formula found in Appendix C. Formulas.

The specific alarm level uses the initial "quick weight" x 0.9 to determine the

alarm level used for scan time calculations. This provides a safety margin in the

calculations.

The final alarm level comparison is based on the more accurate weight

normalized to count time.

NOTE:

The minimum count time calculation is disabled if N*Sigma alarm type is

selected. If N*Sigma alarm type is selected in conjunction with a "0" count time,

the count time will automatically be set to 30 seconds.

Count

Display:

Selects the method used for displaying counts on the screen and in the reports.

The background counts will always be displayed in counts/second. The count

display options only control the display after a scan.

The available options are:

1. Total counts per scan: Displays total counts per scan time.

2. Net counts per scan: Displays counts per scan time above background.

3. Total counts per second: Displays total counts per second.

4. Net counts per second: Displays counts per second above background.

MCA

Data:

MCA data is an optional entry feature.