3. OPERATING INSTRUCTIONS page 17

Model BM-286A

3.4.2. Set-Up Waste Streams:

The following parameters must be set for each waste stream to be used:

WASTE STREAM PARAMETERS

Waste I.D.

Alarm Message

Count Time

Clean Message

Count Display

Maximum Weight

MCA Data

Efficiency

Alarm Type

Alarm Level

A detailed description of each parameter entry follows:

Waste

I.D.:

This parameter provides a description of the waste stream. The operator may

program up to 16 alpha-numeric characters to provide a unique reference for each

waste stream. The existing I.D. will be displayed.

Count

Time:

Scan time, in seconds. Enter a value between 1 and 50,000, or "0" for minimum

count time. If minimum count time is selected, the system will automatically

calculate the minimum time required to scan the item. The following are assumed

for automatic scan time calculation:

bkg = background in counts/second

consig = confidence sigma (fixed at 1.65Õ or 95%)

fasig = false alarm sigma (fixed at 2.33Õ or 1/100)

alev = alarm level in cps/efficiency

efficiency = waste stream efficiency