3. OPERATING INSTRUCTIONS page 21

Model BM-286A

Alarm

Level:

The operator will be prompted for the alarm level (depending on the alarm type

selected above). Enter the alarm level (above background) in the appropriate

units.

The previous value will be displayed, and will only be accepted by entering it

again, or a new value may be entered.

1. DPM:

Radiation alarm level in disintegrations per minute. Enter a value between

10,000 and 1,000,000,000

. This is the level above background that will generate

a radiation alarm when the DPM alarm type is selected.

NOTE:

This value is used only when the Alarm Type is set to DPM.

2. N*Sigma:

Radiation alarm level in N*Sigma (N*1background + background). This is the

count level that will generate a radiation alarm when N*Sigma alarm type is

selected.

NOTE:

This value is used only when the Alarm Type is set to N*Sigma.

3. SpAct: Radiation alarm level, above background, in pCi per gram.